Lablisa® E2(Estradiol) ELISA Kit

Product Information

Lablisa® E2(Estradiol) ELISA Kit

Catalog No : LAB13398 | Pack Size : 48T , 96T

This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with E2. Standards or samples are added to the appropriate micro ELISA plate wells then with a biotin-conjugated antibody specific to E2. Next,Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of E2 in the samples is then determined by comparing the OD of the samples to the standard curve.

Product name:

Lablisa® E2(Estradiol) ELISA Kit
Reactivity:Universal
Alternative Names:Oestradiol; E2; 17β-Estradiol; Estra-1,3,5(10)-triene-3,17β-diol
Assay Type:Competitive
Sensitivity:7.5 pg/mL
Standard:800pg/mL
Range:12.5-800 pg/mL
Sample Type:Serum, Plasma, Other biological fluids,Sample volume: 50μL
Assay Length:1.5h
Research Area:Signaling transduction;Cancer;Metabolism

Standard curve



Concentration (ng/mL)OD
800.000.185
400.000.345
200.000.624
100.000.968
50.001.385
25.001.648
12.501.825
0.002.061


Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.


Recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

MatrixRecovery rangeAverage
serum(n=8)83-97%90%
EDTA plasma(n=8)80-97%88%
Cell culture media (n=8)85-101%93%


Linearity

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Matrix1:21:41:81:16
serum(n=5)85-94%87-96%92-105%89-102%
EDTA plasma(n=5)85-102%85-102%80-92%87-103%
Cell culture media (n=5)89-98%80-97%96-104%85-93%