Lablisa® Human 2,3-DPG(2,3-Diphosphoglycerate) ELISA Kit

Product Information

Lablisa® Human 2,3-DPG(2,3-Diphosphoglycerate) ELISA Kit

Catalog No : LAB0778 | Pack Size : 48T , 96T

The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human 2,3-DPG. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human 2,3-DPG. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human 2,3-DPG, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human 2,3-DPG in the samples is then determined by comparing the OD of the samples to the standard curve.

Product name:Human 2,3-DPG(2,3-Diphosphoglycerate) ELISA Kit
Reactivity:Human
Alternative Names:2,3DPG
Assay Type:Sandwich
Sensitivity:0.14 μmol/mL
Standard:20 μmol/mL
Detection Range:0.32-20 μmol/mL
Sample Type:Serum, plasma, tissue homogenates, cell lysates and other biological fluids
Assay Length:3.5h
Research Area:

Metabolic pathway


Standard curve



Concentration (μmol/mL)ODCorrected OD
20.002.1542.059
10.001.5871.492
5.001.1841.089
2.500.8660.771
1.250.5190.424
0.630.3360.241
0.320.1900.095
0.000.0950.000


Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.


Recovery

Matrices listed below were spiked with certain level of recombinant 2,3-DPG and the recovery rates were calculated by comparing the measured value to the expected amount of 2,3-DPG in samples.

MatrixRecovery rangeAverage
serum(n=5)90-105%97%
EDTA plasma(n=5)80-92%86%
Heparin plasma(n=5)88-102%95%


Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of 2,3-DPG and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Matrix1:21:41:81:16
serum(n=5)85-97%90-99%82-101%93-105%
EDTA plasma(n=5)85-97%87-96%82-95%92-105%
Heparin plasma(n=5)87-96%92-105%89-102%92-101%