Lablisa® FastTrack Rat VEGFR2(Vascular Endothelial Growth Factor Receptor 2) ELISA Kit

Product Information

Lablisa® FastTrack Rat VEGFR2(Vascular Endothelial Growth Factor Receptor 2) ELISA Kit

Catalog No : LAB083FT | Pack Size : 48T , 96T

The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Rat VEGFR2, and the Rat VEGFR2 standard plate wells that pre-coated using protein-related techniques are provided separately. Standard/Sample Diluent Buffer or samples are added to the appropriate microtiter plate wells ,then added a HRP-conjugated antibody specific to Rat VEGFR2. After TMB substrate solution is added, only those wells that contain Rat VEGFR2 and HRP-conjugated antibody will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Rat VEGFR2 in the samples is then determined by comparing the OD of the samples to the standard curve.

Product name:Lablisa® FastTrack Rat VEGFR2(Vascular Endothelial Growth Factor Receptor 2) ELISA Kit
Reactivity:Rat
Alternative Names:CD309; FLK1; VEGFR; KDR; A Type III Receptor Tyrosine Kinase; Kinase Insert Domain Receptor; Kinase Insert Domain Receptor; Fetal Liver Kinase-1
Assay Type:Sandwich
Sensitivity:157.5 pg/mL
Standard:50 ng/mL
Detection Range:0.78-50 ng/mL
Sample Type:Serum, plasma
Assay Length:1.5h
Research Area:Signal transduction;CD & Adhesion molecule;Tumor immunity;

Standard curve



Concentration (ng/mL)ODCorrected OD
50.002.1682.074
25.001.6571.563
12.501.1161.022
6.250.8950.801
3.130.4280.334
1.570.4230.329
0.790.2480.154
0.000.0940.000


Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.


Recovery

Matrices listed below were spiked with certain level of recombinant VEGFR2 and the recovery rates were calculated by comparing the measured value to the expected amount of VEGFR2 in samples.

MatrixRecovery rangeAverage
serum(n=5)79-93%86%
EDTA plasma(n=5)95-107%101%
Heparin plasma(n=5)85-95%90%


Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of VEGFR2 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Matrix1:21:41:81:16
serum(n=5)80-101%87-96%96-105%95-103%
EDTA plasma(n=5)98-105%96-103%87-95%83-92%
Heparin plasma(n=5)95-104%87-96%89-101%97-105%


Manuals on the web are for your reference only, specifications are subject to the delivery manuals.