Lablisa® HVD3(25-Hydroxyvitamin D3) ELISA Kit
Product Information
- Description
- More Infomation
This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with HVD3. Standards or samples are added to the appropriate micro ELISA plate wells then with a biotin-conjugated antibody specific to HVD3. Next,Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of HVD3 in the samples is then determined by comparing the OD of the samples to the standard curve.
Product name: | Lablisa® HVD3(25-Hydroxyvitamin D3) ELISA Kit |
Reactivity: | Universal |
Alternative Names: | 25-HVD3; 25HVD3; Calcifediol; Calcidiol; H-VD3; 25-Hydroxycholecalciferol; 25-Hydroxy Vitamin D3; 25(OH)D |
Assay Type: | Competitive |
Sensitivity: | 0.91 ng/mL |
Standard: | 200 ng/mL |
Range: | 3.13-200 ng/mL |
Sample Type: | Serum, Plasma and Other biological fluids |
Assay Length: | 1.5h |
Research Area: | Reproductive science;Genetic science;Nutrition metabolism;Bone metabolism; |
Standard curve
Concentration (ng/mL) | OD |
---|---|
200.00 | 0.188 |
100.00 | 0.324 |
50.00 | 0.539 |
25.00 | 0.873 |
12.50 | 1.301 |
6.25 | 1.693 |
3.13 | 2.069 |
0.00 | 2.509 |
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.
Recovery
The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
Matrix | Recovery range | Average |
---|---|---|
serum(n=8) | 80-93% | 85% |
EDTA plasma(n=8) | 87-99% | 93% |
Cell culture media (n=8) | 90-103% | 96% |
Linearity
Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 96-101% | 85-92% | 80-96% | 89-101% |
EDTA plasma(n=5) | 87-98% | 87-98% | 92-104% | 82-96% |
Cell culture media (n=5) | 95-102% | 89-97% | 86-102% | 94-106% |