Lablisa® Human IGFBP2(Insulin Like Growth Factor Binding Protein 2) ELISA Kit

Product Information

Lablisa® Human IGFBP2(Insulin Like Growth Factor Binding Protein 2) ELISA Kit

Catalog No : LAB1138 | Pack Size : 48T & 96T 

The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Insulin Like Growth Factor Binding Protein 2(IGFBP2). Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Insulin Like Growth Factor Binding Protein 2(IGFBP2). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Insulin Like Growth Factor Binding Protein 2(IGFBP2), biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Insulin Like Growth Factor Binding Protein 2(IGFBP2) in the samples is then determined by comparing the OD of the samples to the standard curve.

Product name:Human IGFBP2(Insulin Like Growth Factor Binding Protein 2) ELISA Kit
Reactivity:Human
Alternative Names:IBP2; IGF-BP53
Assay Type:Sandwich
Sensitivity:27.2 pg/mL
Standard:4000 pg/mL
Sample type:serum, plasma, urine, tissue homogenates and other biological fluids
Assay length:3.5h
Research Area:Tumor immunity;Endocrinology;

Standard curve



Concentration (ng/mL)ODCorrected OD
4000.002.0961.997
2000.001.6021.503
1000.001.1211.022
500.000.8920.793
250.000.5470.448
125.000.3150.216
62.500.1960.097
0.000.0990.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.


Recovery

Matrices listed below were spiked with certain level of recombinant IGFBP2 and the recovery rates were calculated by comparing the measured value to the expected amount of IGFBP2 in samples.

MatrixRecovery rangeAverage
serum(n=5)78-92%85%
EDTA plasma(n=5)78-90%84%
Heparin plasma(n=5)85-99%92%


Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of IGFBP2 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Matrix1:21:41:81:16
serum(n=5)86-99%93-107%85-94%87-101%
EDTA plasma(n=5)86-97%88-95%92-101%87-98%
Heparin plasma(n=5)85-92%79-96%89-100%85-96%