Lablisa® FastTrack Human PGII(Pepsinogen II) ELISA Kit

Product Information

Lablisa® FastTrack Human PGII(Pepsinogen II) ELISA Kit

Catalog No : LAB010FT | Pack Size : 48T , 96T

The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human PGII, and the Human PGII standard plate wells that pre-coated using protein-related techniques are provided separately. Standard/Sample Diluent Buffer or samples are added to the appropriate microtiter plate wells ,then added a HRP-conjugated antibody specific to Human PGII. After TMB substrate solution is added, only those wells that contain Human PGII and HRP-conjugated antibody will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human PGII in the samples is then determined by comparing the OD of the samples to the standard curve.

Product name:Lablisa® FastTrack Human PGII(Pepsinogen II) ELISA Kit
Reactivity:Human
Alternative Names:PGII
Assay Type:Sandwich
Sensitivity:7 pg/mL
Standard:10 ng/mL
Detection Range:0.16-10 ng/mL
Sample Type:Serum, plasma
Assay Length:1.5h
Research Area:Enzyme & Kinase;Tumor immunity;Gastroenterology;

Standard curve



Concentration (ng/mL)ODCorrected OD
10.002.0581.965
5.001.6281.535
2.501.1091.016
1.250.8210.728
0.630.4740.381
0.320.3610.268
0.160.2280.135
0.000.0930.000


Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.


Recovery

Matrices listed below were spiked with certain level of recombinant PGII and the recovery rates were calculated by comparing the measured value to the expected amount of PGII in samples.

MatrixRecovery rangeAverage
serum(n=5)90-105%97%
EDTA plasma(n=5)88-103%95%
Heparin plasma(n=5)92-106%99%


Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of PGII and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Matrix1:21:41:81:16
serum(n=5)87-102%82-91%86-93%81-94%
EDTA plasma(n=5)91-98%83-101%87-98%95-102%
Heparin plasma(n=5)88-97%78-96%80-99%89-97%


Manuals on the web are for your reference only, specifications are subject to the delivery manuals.