Lablisa® Human SIRT1(Sirtuin 1) ELISA Kit

Product Information

Lablisa® Human SIRT1(Sirtuin 1) ELISA Kit

Catalog No : LAB18393 | Pack Size : 48T , 96T

This ELISA kit uses the Sandwich-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with an antibody specific to Human SIRT1. Standards or samples are added to the micro ELISA plate wells and combined with the specific antibody. Then a biotinylated detection antibody specific for Human SIRT1 and Avidin-Horseradish Peroxidase (HRP) conjugate are added successively to each micro plate well and incubated. Free components are washed away. The substrate solution is added to each well. Only those wells that contain Human SIRT1, biotinylated detection antibody and Avidin-HRP conjugate will appear blue in color. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The OD value is proportional to the concentration of Human SIRT1. You can calculate the concentration of Human SIRT1 in the samples by comparing the OD of the samples to the standard curve.

Product name:

Lablisa® Human SIRT1(Sirtuin 1) ELISA Kit
Reactivity:Human
Alternative Names:SIR2L1, NAD-dependent deacetylase sirtuin-1
Assay Type:Sandwich
Sensitivity:0.19 ng/mL
Standard:20ng/mL
Range:0.32-20ng/mL
Sample Type:Tissue homogenate, Cell lysate;Sample Volume=100μL
Assay Length:3.5h
Research Area:Signal transduction;


Standard curve



Concentration (ng/mL)ODCorrected OD
20.002.336
2.284
10.001.542
1.490
5.000.911
0.859
2.500.4280.376
1.250.2490.197
0.630.149
0.097
0.320.1020.050
0.000.052
0.000


Precision

Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.

Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.


Recovery

The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.

MatrixRecovery rangeAverage
Serum (n=8)
84-96%90%
EDTA plasma (n=8)
87-95%91%
Cell culture media (n=8)
95-107%101%


Linearity

Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.

Matrix1:21:41:81:16
Serum (n=5)
85-97%95-107%92-103%81-99%
EDTA plasma (n=5)
80-96%93-105%83-97%95-104%
Cell culture media(n=5)82-93%81-96%80-91%83-94%


Manuals on the web are for your reference only, specifications are subject to the delivery manuals.