This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with PGE2. Standards or samples are added to the appropriate micro ELISA plate wells then with a biotin-conjugated antibody specific to PGE2. Next,Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of PGE2 in the samples is then determined by comparing the OD of the samples to the standard curve.
Product name: | Lablisa® PGE2(Prostaglandin E2) ELISA Kit |
Reactivity: | Universal |
Alternative Names: | PG-E2 |
Assay Type: | Competitive |
Sensitivity: | 0.97 pg/mL |
Standard: | 200 pg/mL |
Range: | 3.13-200 pg/mL |
Sample Type: | Serum, Plasma and Other biological fluids |
Assay Length: | 1.5h |
Research Area: | Metabolic pathway;Infection immunity;Endocrinology;Hormone metabolism; |
Standard curve
Concentration (ng/mL) | OD |
---|---|
200.00 | 0.186 |
100.00 | 0.322 |
50.00 | 0.537 |
25.00 | 0.871 |
12.50 | 1.299 |
6.25 | 1.691 |
3.13 | 2.067 |
0.00 | 2.507 |
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.
Recovery
The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
Matrix | Recovery range | Average |
---|---|---|
serum(n=8) | 93-107% | 100% |
EDTA plasma(n=8) | 87-99% | 93% |
Cell culture media (n=8) | 80-93% | 86% |
Linearity
Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 91-99% | 83-89% | 88-94% | 87-93% |
EDTA plasma(n=5) | 92-101% | 92-101% | 82-98% | 89-97% |
Cell culture media (n=5) | 85-93% | 86-93% | 93-101% | 86-97% |