Lablisa® Chicken IgY(Immunoglobulin of Yolk) ELISA Kit

Product Information

Lablisa® Chicken IgY(Immunoglobulin of Yolk) ELISA Kit

Catalog No : LAB9182 | Pack Size : 48T , 96T

This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with Chicken IgY protein. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Chicken IgY. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Chicken IgY in the samples is then determined by comparing the OD of the samples to the standard curve.

Product name:Chicken IgY(Immunoglobulin of Yolk) ELISA Kit
Reactivity:Chicken
Alternative Names:Immunoglobulin Yolk
Assay Type:Competitive Inhibition
Sensitivity:1.85 μg/mL
Standard:400 μg/mL
Detection Range:6.25-400 μg/mL
Sample Type:serum, plasma and other biological fluids
Assay Length:2h
Research Area:Infection immunity;Immune molecule;Hematology;


Standard curve



Concentration (μg/mL)ODCorrected OD
400.000.199
200.000.397
100.000.577
50.000.827
25.001.289
12.501.537
6.251.861
0.00
2.101


Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.


Recovery

Matrices listed below were spiked with certain level of recombinant IgY and the recovery rates were calculated by comparing the measured value to the expected amount of IgY in samples.

MatrixRecovery rangeAverage
serum(n=5)83-98%90%
EDTA plasma(n=5)92-107%99%
Heparin plasma(n=5)87-99%93%


Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of IgY and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Matrix1:21:41:81:16
serum(n=5)87-98%95-102%84-99%81-95%
EDTA plasma(n=5)88-97%93-103%94-106%91-101%
Heparin plasma(n=5)80-92%86-102%87-103%85-101%