Lablisa® FastTrack Mouse IL-1β(Interleukin 1 Beta) ELISA Kit

Product Information

Lablisa® FastTrack Mouse IL-1β(Interleukin 1 Beta) ELISA Kit

Catalog No : LAB13643FT | Pack Size : 48T , 96T

This assay has high sensitivity and excellent specificity for detection of Mouse IL-1β. No significant cross-reactivity or interference between Mouse IL-1β and analogues was observed.

Product name:Lablisa® FastTrack Mouse IL-1β(Interleukin 1 Beta) ELISA Kit
Reactivity:Mouse
Alternative Names:IL1B, IL1-BETA, IL1F2, catabolin, IL1b, IL-1B, IL-1b
Assay Type:Sandwich
Sensitivity:2.89 pg/mL
Standard:400 pg/mL
Range:62.5-400 pg/mL
Sample Type:Serum, Plasma, Tissue homogenate and Other biological samples ;Sample Volume=50μL
Assay Length:1.5H
Research Area:
Cancer, Immunology, Metabolism

Standard curve



Concentration (ng/mL)ODCorrected OD
4002.5412.448
2001.7351.642
1000.9810.888
500.6110.518
250.3560.263
12.50.2320.139
6.250.1510.058
0.000.0930.00


Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.


Recovery

Matrices listed below were spiked with certain level of recombinant Mouse IL-1β and the recovery rates were calculated by comparing the measured value to the expected amount of Mouse IL-1β in samples.

MatrixRecovery rangeAverage
serum(n=8)86-99%92%
EDTA plasma(n=8)80-93%86%
Heparin plasma(n=8)90-105%97%


Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Mouse IL-1β and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Matrix1:21:41:81:16
serum(n=8)85-97%86-102%87-99%99-105%
EDTA plasma(n=8)95-104%82-94%86-96%81-97%
Heparin plasma(n=8)92-107%83-99%80-96%91-101%