Lablisa® ACH(Acetylcholine) ELISA Kit
Product Information
- Description
- More Infomation
This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with ACH. Standards or samples are added to the appropriate micro ELISA plate wells then with a biotin-conjugated antibody specific to ACH. Next,Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of ACH in the samples is then determined by comparing the OD of the samples to the standard curve.
Product name: | Lablisa® ACH(Acetylcholine) ELISA Kit |
| Reactivity: | Universal |
| Alternative Names: | Acetylcholine |
| Assay Type: | Competitive |
| Sensitivity: | 5.31 pg/mL |
| Standard: | 1000 pg/mL |
| Range: | 15.63-1000 pg/mL |
| Sample Type: | Serum, Plasma and Other biological fluids |
| Assay Length: | 1.5h |
| Research Area: | Signal transduction;Infection immunity;Neuro science; |
Standard curve

| Concentration (ng/mL) | OD |
|---|---|
| 1000.00 | 0.279 |
| 500.00 | 0.379 |
| 250.00 | 0.573 |
| 125.00 | 0.834 |
| 62.50 | 1.129 |
| 31.25 | 1.524 |
| 15.63 | 1.838 |
| 0.00 | 2.502 |
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.
Recovery
The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=8) | 80-90% | 84% |
| EDTA plasma(n=8) | 85-97% | 91% |
| Cell culture media (n=8) | 80-93% | 86% |
Linearity
Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 85-96% | 87-101% | 89-102% | 85-98% |
| EDTA plasma(n=5) | 86-101% | 86-101% | 86-101% | 87-96% |
| Cell culture media (n=5) | 92-105% | 89-102% | 92-101% | 88-102% |