Lablisa® FastTrack Human IL-2sRa/CD25(Soluble Interleukin-2 Receptor alpha chain) ELISA Kit

Product Information

Lablisa® FastTrack Human IL-2sRa/CD25(Soluble Interleukin-2 Receptor alpha chain) ELISA Kit

Catalog No : LAB087FT | Pack Size : 48T , 96T

The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human IL-2Ra/CD25, and the Human IL-2Ra/CD25 standard plate wells that pre-coated using protein-related techniques are provided separately. Standard/Sample Diluent Buffer or samples are added to the appropriate microtiter plate wells ,then added a HRP-conjugated antibody specific to Human IL-2Ra/CD25. After TMB substrate solution is added, only those wells that contain Human IL-2Ra/CD25 and HRP-conjugated antibody will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human IL-2Ra/CD25 in the samples is then determined by comparing the OD of the samples to the standard curve.

Product name:Lablisa® FastTrack Human IL-2sRa/CD25(Soluble Interleukin-2 Receptor alpha chain) ELISA Kit
Reactivity:Human
Alternative Names:IL-2RA;CD25; IL2R-a; IL2-RA; IDDM10; IL2R; TCGFR; p55; TAC antigen; IL2Ra; Interleukin 2 Receptor Alpha
Assay Type:Sandwich
Sensitivity:37.66 pg/mL
Standard:6000 pg/mL
Detection Range:93.75-6000 pg/mL
Sample Type:Serum, plasma
Assay Length:1.5h
Research Area:Signal transduction;CD & Adhesion molecule;Tumor immunity;Infection immunity;

Standard curve



Concentration (pg/mL)ODCorrected OD
6000.002.1892.088
3000.001.5821.481
1500.001.2431.142
750.000.9350.834
375.000.5030.402
187.500.2850.184
93.750.2190.118
0.000.1010.000


Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.


Recovery

Matrices listed below were spiked with certain level of recombinant IL-2Ra/CD25 and the recovery rates were calculated by comparing the measured value to the expected amount of IL-2Ra/CD25 in samples.

MatrixRecovery rangeAverage
serum(n=5)81-97%89%
EDTA plasma(n=5)85-99%92%
Heparin plasma(n=5)90-105%97%


Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of IL-2Ra/CD25 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Matrix1:21:41:81:16
serum(n=5)79-96%85-99%79-93%89-97%
EDTA plasma(n=5)88-105%86-92%81-93%79-97%
Heparin plasma(n=5)85-102%87-96%92-101%89-104%