Lablisa® Human D2D(D-Dimer) ELISA Kit

Product Information

Lablisa® Human D2D(D-Dimer) ELISA Kit

Catalog No : LAB1213| Pack Size : 48T , 96T

The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human D2D. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human D2D. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human D2D, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human D2D in the samples is then determined by comparing the OD of the samples to the standard curve.

Product name:Human D2D(D-Dimer) ELISA Kit
Reactivity:Human
Alternative Names:D 2 Dimer
Assay Type:Sandwich
Sensitivity:26.7 ng/mL
Standard:5000 ng/mL
Detection Range:78.13-5000 ng/mL
Sample Type:plasma
Assay Length:3.5h
Research Area:Hematology;Cardiovascular biology;


Standard curve



Concentration (ng/mL)ODCorrected OD
5000.002.1312.048
2500.001.6681.585
1250.001.1831.100
625.000.8920.809
312.500.5890.506
156.250.3650.282
78.130.2570.174
0.000.0830.000


Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.


Recovery

Matrices listed below were spiked with certain level of recombinant D2D and the recovery rates were calculated by comparing the measured value to the expected amount of D2D in samples.

MatrixRecovery rangeAverage
serum(n=5)81-95%88%
EDTA plasma(n=5)83-95%89%
Heparin plasma(n=5)87-97%90%


Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of D2D and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Matrix1:21:41:81:16
serum(n=5)98-107%93-102%80-93%91-100%
EDTA plasma(n=5)85-102%83-96%95-102%87-98%
Heparin plasma(n=5)87-94%85-96%89-100%85-96%


Manuals on the web are for your reference only, specifications are subject to the delivery manuals.