Lablisa® Human PAI1(Plasminogen Activator Inhibitor 1) ELISA Kit

Product Information

Lablisa® Human PAI1(Plasminogen Activator Inhibitor 1) ELISA Kit

Catalog No : LAB1250 | Pack Size : 48T & 96T 

The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Plasminogen Activator Inhibitor 1(PAI1). Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Plasminogen Activator Inhibitor 1(PAI1). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Plasminogen Activator Inhibitor 1(PAI1), biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Plasminogen Activator Inhibitor 1(PAI1) in the samples is then determined by comparing the OD of the samples to the standard curve.

Product name:Human PAI1(Plasminogen Activator Inhibitor 1) ELISA Kit
Reactivity:Human
Alternative Names:SERPINE1; PLANH1; Serpin Peptidase Inhibitor Clade E Member 1; Nexin,Plasminogen Activator Inhibitor Type 1 Serpin E1; Endothelial plasminogen activator inhibitor
Assay Type:Sandwich
Sensitivity:5.1 pg/mL
Standard:800 pg/mL
Sample type:plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay length:3.5h
Research Area:Metabolic pathway;Hematology;

Standard curve


Concentration (ng/mL)ODCorrected OD
800.002.0541.971
400.001.7821.699
200.001.1351.052
100.000.8920.809
50.000.5620.479
25.000.3210.238
12.500.2270.144
0.000.0830.000


Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.


Recovery

Matrices listed below were spiked with certain level of recombinant PAI1 and the recovery rates were calculated by comparing the measured value to the expected amount of PAI1 in samples.

MatrixRecovery rangeAverage
serum(n=5)80-95%87%
EDTA plasma(n=5)83-97%90%
Heparin plasma(n=5)78-92%85%


Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of PAI1 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Matrix1:21:41:81:16
serum(n=5)87-98%92-104%82-96%95-102%
EDTA plasma(n=5)87-98%92-105%83-92%89-98%
Heparin plasma(n=5)83-97%86-97%90-98%80-93%


Manuals on the web are for your reference only, specifications are subject to the delivery manuals.