Lablisa® Mouse NT3(Neurotrophin 3) ELISA Kit

Product Information

Lablisa® Mouse NT3(Neurotrophin 3) ELISA Kit

Catalog No : LAB1175| Pack Size : 48T , 96T

The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Mouse NT3. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Mouse NT3. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Mouse NT3, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Mouse NT3 in the samples is then determined by comparing the OD of the samples to the standard curve.

Product name:Mouse NT3(Neurotrophin 3) ELISA Kit
Reactivity:Mouse
Alternative Names:NTF3; HDNF; NGF2; Nerve Growth Factor 2; Neurotrophic Factor
Assay Type:Sandwich
Sensitivity:30 pg/mL
Standard:5000 pg/mL
Detection Range:78.13-5000 pg/mL
Sample Type:Serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Length:3.5h
Research Area:Cytokine;Neuro science;


Standard curve



Concentration (pg/mL)ODCorrected OD
5000.001.9411.859
2500.001.4941.412
1250.001.2431.161
625.000.8390.757
312.500.5390.457
156.250.3880.306
78.130.2110.129
0.000.0820.000


Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.


Recovery

Matrices listed below were spiked with certain level of recombinant NT3 and the recovery rates were calculated by comparing the measured value to the expected amount of NT3 in samples.

MatrixRecovery rangeAverage
serum(n=5)80-97%88%
EDTA plasma(n=5)82-98%90%
Heparin plasma(n=5)82-94%88%


Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of NT3 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Matrix1:21:41:81:16
serum(n=5)98-105%92-101%85-105%87-98%
EDTA plasma(n=5)85-92%79-96%89-100%85-96%
Heparin plasma(n=5)86-93%93-101%85-98%90-101%