Lablisa® Human T4(Thyroxine) ELISA Kit

Product Information

Lablisa® Human T4(Thyroxine) ELISA Kit

Catalog No : LAB0916 | Pack Size : 48T , 96T

This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with Human T4 protein. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human T4. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human T4 in the samples is then determined by comparing the OD of the samples to the standard curve.

Product name:Human T4(Thyroxine) ELISA Kit
Reactivity:Human
Alternative Names:3,5,3',5'-Tetraiodothyronine; L-Thyroxine; Levothyroxine
Assay Type:Competitive Inhibition
Sensitivity:1.24 ng/mL
Standard:300 ng/mL
Detection Range:4.69-300 ng/mL
Sample Type:serum, plasma and other biological fluids
Assay Length:2h
Research Area:Endocrinology;Hormone metabolism;


Standard curve



Concentration (ng/mL)ODCorrected OD
300.000.182
150.000.357
75.000.567
37.500.856
18.751.225
9.381.579
4.691.861
0.00
2.197


Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.


Recovery

Matrices listed below were spiked with certain level of recombinant T4 and the recovery rates were calculated by comparing the measured value to the expected amount of T4 in samples.

MatrixRecovery rangeAverage
serum(n=5)82-98%90%
EDTA plasma(n=5)84-98%91%
Heparin plasma(n=5)80-96%88%


Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of T4 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Matrix1:21:41:81:16
serum(n=5)95-104%87-96%88-101%82-95%
EDTA plasma(n=5)79-92%88-96%77-91%85-93%
Heparin plasma(n=5)86-97%93-102%87-98%83-97%