Lablisa® Human MMP8(Matrix Metalloproteinase 8) ELISA Kit

Product Information

Lablisa® Human MMP8(Matrix Metalloproteinase 8) ELISA Kit

Catalog No : LAB1172 | Pack Size : 48T , 96T

The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human MMP8. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human MMP8. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human MMP8, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human MMP8 in the samples is then determined by comparing the OD of the samples to the standard curve.

Product name:
Human MMP8(Matrix Metalloproteinase 8) ELISA Kit
Reactivity:Human
Alternative Names:CLG2; HNC; PMNL-CL; Collagenase 2; Neutrophil Collagenase; PMNL collagenase
Assay Type:Sandwich
Sensitivity:28 pg/mL
Standard:5000 pg/mL
Detection Range:78.13-5000 pg/mL
Sample Type:serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Length:3.5h
Research Area:Enzyme & Kinase;Infection immunity;Cardiovascular biology;


Standard curve



Concentration (ng/mL)ODCorrected OD
5000.002.0271.954
2500.001.4971.424
1250.001.0690.996
625.000.8830.810
312.500.5590.486
156.250.3030.230
78.130.2320.159
0.000.0730.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.


Recovery

Matrices listed below were spiked with certain level of recombinant MMP8 and the recovery rates were calculated by comparing the measured value to the expected amount of MMP8 in samples.

MatrixRecovery rangeAverage
serum(n=5)90-105%97%
EDTA plasma(n=5)80-95%87%
Heparin plasma(n=5)86-99%92%


Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of MMP8 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.