Lablisa® SPH(Sphingomyelin) ELISA Kit
Product Information
- Description
- More Infomation
This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with SPH. Standards or samples are added to the appropriate micro ELISA plate wells then with a biotin-conjugated antibody specific to SPH. Next,Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of SPH in the samples is then determined by comparing the OD of the samples to the standard curve.
Product name: | Lablisa® SPH(Sphingomyelin) ELISA Kit |
| Reactivity: | Universal |
| Alternative Names: | SM |
| Assay Type: | Competitive |
| Sensitivity: | 1.85 μg/mL |
| Standard: | 400 μg/mL |
| Range: | 6.25-400 μg/mL |
| Sample Type: | Serum, Plasma and Other biological fluids |
| Assay Length: | 1.5h |
| Research Area: | Signal transduction;Neuro science; |
Standard curve

| Concentration (ng/mL) | OD |
|---|---|
| 400.00 | 0.254 |
| 200.00 | 0.417 |
| 100.00 | 0.657 |
| 50.00 | 0.974 |
| 25.00 | 1.267 |
| 12.50 | 1.587 |
| 6.25 | 1.857 |
| 0.00 | 2.679 |
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.
Recovery
The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=8) | 80-93% | 85% |
| EDTA plasma(n=8) | 87-99% | 93% |
| Cell culture media (n=8) | 90-103% | 96% |
Linearity
Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 87-97% | 86-92% | 89-103% | 89-98% |
| EDTA plasma(n=5) | 80-85% | 80-85% | 87-96% | 82-95% |
| Cell culture media (n=5) | 89-103% | 95-102% | 96-104% | 87-98% |