Lablisa® Rat TRPV4(Transient Receptor Potential Cation Channel Subfamily V, Member 4) ELISA Kit

Product Information

Lablisa® Rat TRPV4(Transient Receptor Potential Cation Channel Subfamily V, Member 4) ELISA Kit

Catalog No : LAB0818 | Pack Size : 48T , 96T

The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Rat TRPV4. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Rat TRPV4. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Rat TRPV4, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Rat TRPV4 in the samples is then determined by comparing the OD of the samples to the standard curve.

Product name:Rat TRPV4(Transient Receptor Potential Cation Channel Subfamily V, Member 4) ELISA Kit
Reactivity:Rat
Alternative Names:SMAL; VRL2; BCYM3; CMT2C; SPSMA; TRP12; VROAC; HMSN2C; OTRPC4; SSQTL1
Assay Type:Sandwich
Sensitivity:18.87 pg/mL
Standard:3000 pg/mL
Detection Range:46.88-3000 pg/mL
Sample Type:Tissue homogenates, cell lysates and other biological fluids.
Assay Length:3.5h
Research Area:Signal transduction;

Standard curve



Concentration (pg/mL)ODCorrected OD
3000.002.1312.036
1500.001.6681.573
750.001.1201.025
375.000.8060.711
187.500.5590.464
93.750.3360.241
46.880.2360.141
0.000.0950.000


Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.


Recovery

Matrices listed below were spiked with certain level of recombinant TRPV4 and the recovery rates were calculated by comparing the measured value to the expected amount of TRPV4 in samples.

MatrixRecovery rangeAverage
serum(n=5)87-99%92%
EDTA plasma(n=5)79-97%88%
Heparin plasma(n=5)80-92%86%


Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of TRPV4 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Matrix1:21:41:81:16
serum(n=5)95-103%95-103%84-98%93-102%
EDTA plasma(n=5)87-98%93-105%89-97%85-97%
Heparin plasma(n=5)85-92%82-96%87-105%83-92%