Lablisa® PREG(Pregnenolone) ELISA Kit
Product Information
- Description
- More Infomation
This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with P5. Standards or samples are added to the appropriate micro ELISA plate wells then with a biotin-conjugated antibody specific to P5. Next,Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of P5 in the samples is then determined by comparing the OD of the samples to the standard curve.
Product name: | Lablisa® PREG(Pregnenolone) ELISA Kit |
| Reactivity: | Universal |
| Alternative Names: | Pregnenolone |
| Assay Type: | Competitive |
| Sensitivity: | 0.47 ng/mL |
| Standard: | 50 ng/mL |
| Range: | 0.78-50 ng/mL |
| Sample Type: | Serum, Plasma and Other biological fluids |
| Assay Length: | 1.5h |
| Research Area: | Reproductive science |
Standard curve

| Concentration (ng/mL) | OD |
|---|---|
| 50.00 | 0.273 |
| 25.00 | 0.373 |
| 12.50 | 0.567 |
| 6.25 | 0.828 |
| 3.13 | 1.123 |
| 1.56 | 1.518 |
| 0.78 | 1.832 |
| 0.00 | 2.496 |
Precision
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.
Recovery
The recovery of spiked at three different levels in samples throughout the range of the assay was evaluated in various matrices.
| Matrix | Recovery range | Average |
|---|---|---|
| serum(n=8) | 90-107% | 98% |
| EDTA plasma(n=8) | 87-99% | 93% |
| Cell culture media (n=8) | 82-98% | 90% |
Linearity
Samples were spiked with high concentrations of target proteins and diluted with Reference Standard & Sample Diluent to produce samples with values within the range of the assay.
| Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
|---|---|---|---|---|
| serum(n=5) | 93-102% | 90-99% | 95-106% | 87-101% |
| EDTA plasma(n=5) | 85-100% | 85-100% | 84-98% | 95-101% |
| Cell culture media (n=5) | 88-102% | 86-93% | 91-104% | 87-98% |