Lablisa® FastTrack Cort(Corticosterone) ELISA Kit

Product Information

Lablisa® FastTrack Cort(Corticosterone) ELISA Kit

Catalog No : LAB13653FT | Pack Size : 48T , 96T

This assay has high sensitivity and excellent specificity for detection of Cort. No significant cross-reactivity or interference between Cort and analogues was observed.

Product name:

Lablisa® FastTrack Cort(Corticosterone) ELISA Kit
Reactivity:Universal
Alternative Names:Corticosrone
Assay Type:Competitive
Sensitivity:0.51 ng/mL
Standard:100 ng/mL
Range:1.57-100 ng/mL
Sample Type:Serum, Plasma, urine and Other biological fluids; Sample Volume=50μL
Assay Length:0.75H
Research Area:Infection immunity;Endocrinology;Hormone metabolism;

Standard curve



Concentration (ng/mL)OD
1000.274
500.374
250568
12.50.829
6.251.124
3.131.519
1.561.833
0.002.497


Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.


Recovery

Matrices listed below were spiked with certain level of recombinant Cort and the recovery rates were calculated by comparing the measured value to the expected amount of Cort in samples.

MatrixRecovery rangeAverage
serum(n=8)86-99%92%
EDTA plasma(n=8)82-95%88%
Heparin plasma(n=8)80-95%87%


Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of Cort and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Matrix1:21:41:81:16
serum(n=8)85-105%87-98%91-102%86-102%
EDTA plasma(n=8)86-93%86-93%93-101%86-97%
Heparin plasma(n=8)90-101%80-101%87-96%96-105%


Manuals on the web are for your reference only, specifications are subject to the delivery manuals.