Lablisa® FastTrack Mouse 25-OH-D(25 Hydroxy Vitamin D) ELISA Kit
Product Information
- Description
- More Infomation
This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with Mouse 25-OH-D antigen, and the Mouse 25-OH-D standard plate wells that pre-coated using protein-related techniques are provided separately. Standard/Sample Diluent Buffer or samples are added to the appropriate microtiter plate wells ,then added a HRP-conjugated antibody specific to Mouse 25-OH-D to each microplate well and incubated . After TMB substrate solution is added, the enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ±10nm. The concentration of Mouse 25-OH-D in the samples is then determined by comparing the OD of the samples to the standard curve.
Product name: | Lablisa® FastTrack Mouse 25-OH-D(25 Hydroxy Vitamin D) ELISA Kit |
Reactivity: | Mouse |
Alternative Names: | 25 Hydroxy Vitamin D; |
Assay Type: | Competitive Inhibition |
Sensitivity: | 7.7 ng/mL |
Standard: | 5000 ng/mL |
Detection Range: | 78.13-5000 ng/mL |
Sample Type: | Serum, plasma |
Assay Length: | 1.5h |
Research Area: | Reproductive science;Genetic science;Nutrition metabolism;Bone metabolism; |
Standard curve
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
5000.00 | 0.175 | |
2500.00 | 0.352 | |
1250.00 | 0.577 | |
625.00 | 0.833 | |
312.50 | 1.152 | |
156.25 | 1.565 | |
78.13 | 1.789 | |
0.00 | 2.132 |
Precision
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant 25-OH-D and the recovery rates were calculated by comparing the measured value to the expected amount of 25-OH-D in samples.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 95-107% | 101% |
EDTA plasma(n=5) | 82-95% | 88% |
Heparin plasma(n=5) | 78-96% | 87% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of 25-OH-D and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 93-102% | 81-96% | 93-106% | 82-95% |
EDTA plasma(n=5) | 86-97% | 83-96% | 91-103% | 96-105% |
Heparin plasma(n=5) | 85-100% | 80-97% | 81-92% | 87-103% |
Manuals on the web are for your reference only, specifications are subject to the delivery manuals.