Lablisa® Mouse ERK2(Extracellular Signal Regulated Kinase 2) ELISA Kit

Product Information

Lablisa® Mouse ERK2(Extracellular Signal Regulated Kinase 2) ELISA Kit

Catalog No : LAB0945 | Pack Size : 48T , 96T

The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Mouse ERK2. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Mouse ERK2. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Mouse ERK2, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Mouse ERK2 in the samples is then determined by comparing the OD of the samples to the standard curve.

Product name:Mouse ERK2(Extracellular Signal Regulated Kinase 2) ELISA Kit
Reactivity:Mouse
Alternative Names:MAPK1; P38; P40; ERK2; ERT1; MAPK2; P42MAPK; P42-MAPK; PRKM1; PRKM2; P41; P41mapk; P41-mapk; Mitogen-Activated Protein Kinase 1; Mitogen-activated protein kinase 2; ERK1/2
Assay Type:Sandwich
Sensitivity:13.4 pg/mL
Standard:2000 pg/mL
Detection Range:31.25-2000 pg/mL
Sample Type:Serum, plasma, tissue homogenates and other biological fluids
Assay Length:3.5h
Research Area:Signal transduction;Enzyme & Kinase;Tumor immunity;


Standard curve



Concentration (pg/mL)ODCorrected OD
2000.002.0431.956
1000.001.5431.456
500.001.1931.106
250.000.9330.846
125.000.5410.454
62.500.3230.236
31.250.2260.139
0.000.0870.000


Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.


Recovery

Matrices listed below were spiked with certain level of recombinant ERK2 and the recovery rates were calculated by comparing the measured value to the expected amount of ERK2 in samples.

MatrixRecovery rangeAverage
serum(n=5)84-96%90%
EDTA plasma(n=5)78-94%86%
Heparin plasma(n=5)87-95%91%


Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of ERK2 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Matrix1:21:41:81:16
serum(n=5)87-98%92-101%82-98%89-102%
EDTA plasma(n=5)83-96%79-93%81-93%86-98%
Heparin plasma(n=5)92-101%88-102%87-102%93-102%