Lablisa® Human FGF2(Fibroblast Growth Factor 2, Basic) ELISA Kit

Product Information

Lablisa® Human FGF2(Fibroblast Growth Factor 2, Basic) ELISA Kit

Catalog No : LAB1258 | Pack Size : 48T & 96T 

The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Fibroblast Growth Factor 2, Basic(FGF2). Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Fibroblast Growth Factor 2, Basic(FGF2). Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Fibroblast Growth Factor 2, Basic(FGF2), biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm±10nm. The concentration of Fibroblast Growth Factor 2, Basic(FGF2) in the samples is then determined by comparing the OD of the samples to the standard curve.

Product name:Human FGF2(Fibroblast Growth Factor 2, Basic) ELISA Kit
Reactivity:Human
Alternative Names:B-FGF; BFGF; FGFB; HBGH-2; Basic Fibroblast Growth Factor; Heparin-binding growth factor 2
Assay Type:Sandwich
Sensitivity:5.16 pg/mL
Standard:1000 pg/mL
Sample type:serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay length:3.5h
Research Area:Cytokine;Tumor immunity;Infection immunity;

Standard curve



Concentration (ng/mL)ODCorrected OD
1000.001.9521.870
500.001.6471.565
250.001.1141.032
125.000.7820.700
62.500.6320.550
31.250.2520.170
15.630.2250.143
0.000.0820.000

Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.


Recovery

Matrices listed below were spiked with certain level of recombinant FGF2 and the recovery rates were calculated by comparing the measured value to the expected amount of FGF2 in samples.

MatrixRecovery rangeAverage
serum(n=5)82-98%90%
EDTA plasma(n=5)84-98%91%
Heparin plasma(n=5)80-96%88%


Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of FGF2 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix1:21:41:81:16
serum(n=5)79-95%96-105%85-96%90-99%
EDTA plasma(n=5)85-97%93-101%97-103%91-99%
Heparin plasma(n=5)90-98%85-92%83-96%93-102%