Lablisa® Human KLK10(Kallikrein 10) ELISA Kit

Product Information

Lablisa® Human KLK10(Kallikrein 10) ELISA Kit

Catalog No : LAB1110 | Pack Size : 48T , 96T

This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with Human KLK10 protein. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human KLK10. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human KLK10 in the samples is then determined by comparing the OD of the samples to the standard curve.

Product name:Human KLK10(Kallikrein 10) ELISA Kit
Reactivity:Human
Alternative Names:NES1; PRSSL1; Kallikrein-Related Peptidase 10; Normal epithelial cell-specific 1; Protease serine-like 1
Assay Type:Competitive Inhibition
Sensitivity:48.3 pg/mL
Standard:10000 pg/mL
Detection Range:156.25-10000 pg/mL
Sample Type:Serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Length:2h
Research Area:Enzyme & Kinase;Tumor immunity;Reproductive science;


Standard curve



Concentration (pg/mL)ODCorrected OD
10000.000.199
5000.000.385
2500.000.596
1250.000.878
625.001.169
312.501.537
156.251.715
0.00
2.213


Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.


Recovery

Matrices listed below were spiked with certain level of recombinant KLK10 and the recovery rates were calculated by comparing the measured value to the expected amount of KLK10 in samples.

MatrixRecovery rangeAverage
serum(n=5)85-95%90%
EDTA plasma(n=5)79-93%86%
Heparin plasma(n=5)95-107%101%


Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of KLK10 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Matrix1:21:41:81:16
serum(n=5)87-102%85-94%87-101%89-99%
EDTA plasma(n=5)85-101%78-96%79-98%93-102%
Heparin plasma(n=5)92-101%82-98%89-97%85-93%