Lablisa® Mouse ADP(Adiponectin) ELISA Kit

Product Information

Lablisa® Mouse ADP(Adiponectin) ELISA Kit

Catalog No : LAB1004 | Pack Size : 48T , 96T

The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Mouse ADP. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Mouse ADP. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Mouse ADP, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Mouse ADP in the samples is then determined by comparing the OD of the samples to the standard curve.

Product name:Mouse ADP(Adiponectin) ELISA Kit
Reactivity:Mouse
Alternative Names:GBP28; ApM1; AdipoQ; Acrp30; ACDC; APM1; C1Q And Collagen Domain Containing; Adipocyte Complement-Related Protein Of 30 KDa; Adipose Most Abundant Gene Transcript 1
Assay Type:Sandwich
Sensitivity:14.2 pg/mL
Standard:2000 pg/mL
Detection Range:31.25-2000 pg/mL
Sample Type:Serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Length:3.5h
Research Area:Metabolic pathway;Infection immunity;Cardiovascular biology;


Standard curve



Concentration (pg/mL)ODCorrected OD
2000.002.1052.034
1000.001.7001.629
500.001.1431.072
250.000.8380.767
125.000.4960.425
62.500.3520.281
31.250.1520.081
0.000.0710.000


Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.


Recovery

Matrices listed below were spiked with certain level of recombinant ADP and the recovery rates were calculated by comparing the measured value to the expected amount of ADP in samples.

MatrixRecovery rangeAverage
serum(n=5)85-97%91%
EDTA plasma(n=5)78-92%85%
Heparin plasma(n=5)92-105%98%


Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of ADP and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Matrix1:21:41:81:16
serum(n=5)95-104%93-102%79-94%87-98%
EDTA plasma(n=5)86-93%91-105%87-98%89-94%
Heparin plasma(n=5)91-99%83-89%88-104%97-103%