Lablisa® FastTrack Rat 25-ONH-D(25 Hydroxy Vitamin D) ELISA Kit

Product Information

Lablisa® FastTrack Rat 25-ONH-D(25 Hydroxy Vitamin D) ELISA Kit

Catalog No : LAB053FT | Pack Size : 48T , 96T

This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with Rat 25-OH-D antigen, and the Rat 25-OH-D standard plate wells that pre-coated using protein-related techniques are provided separately. Standard/Sample Diluent Buffer or samples are added to the appropriate microtiter plate wells ,then added a HRP-conjugated antibody specific to Rat 25-OH-D to each microplate well and incubated . After TMB substrate solution is added, the enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ±10nm. The concentration of Rat 25-OH-D in the samples is then determined by comparing the OD of the samples to the standard curve.

Product name:Lablisa® FastTrack Rat 25-ONH-D(25 Hydroxy Vitamin D) ELISA Kit
Reactivity:Rat
Alternative Names:25 Hydroxy Vitamin D;
Assay Type:Competitive Inhibition
Sensitivity:7.7 ng/mL
Standard:5000 ng/mL
Detection Range:78.13-5000 ng/mL
Sample Type:Serum, plasma
Assay Length:1.5h
Research Area:Reproductive science;Genetic science;Nutrition metabolism;Bone metabolism;

Standard curve



Concentration (ng/mL)ODCorrected OD
5000.000.189
2500.000.397
1250.000.583
625.000.833
312.501.157
156.251.583
78.131.715
0.002.211


Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.


Recovery

Matrices listed below were spiked with certain level of recombinant 25-OH-D and the recovery rates were calculated by comparing the measured value to the expected amount of 25-OH-D in samples.

MatrixRecovery rangeAverage
serum(n=5)85-97%91%
EDTA plasma(n=5)91-105%98%
Heparin plasma(n=5)95-107%101%


Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of 25-OH-D and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Matrix1:21:41:81:16
serum(n=5)85-107%84-98%95-103%86-98%
EDTA plasma(n=5)86-93%91-105%87-98%99-102%
Heparin plasma(n=5)85-97%87-96%86-98%91-103%


Manuals on the web are for your reference only, specifications are subject to the delivery manuals.