Lablisa® Human LIF(Leukemia Inhibitory Factor) ELISA Kit


Product Information
- Description
- More Infomation
The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human LIF. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human LIF. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human LIF, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human LIF in the samples is then determined by comparing the OD of the samples to the standard curve.
Product name: | Human LIF(Leukemia Inhibitory Factor) ELISA Kit |
Reactivity: | Human |
Alternative Names: | CDF; D-FACTOR; HILDA; MLPLI; Cholinergic Differentiation Factor; Differentiation-Stimulating Factor; Melanoma-Derived LPL Inhibitor; Emfilermin |
Assay Type: | Sandwich |
Sensitivity: | 47 pg/mL |
Standard: | 2000 pg/mL |
Detection Range: | 31.25-2000 pg/mL |
Sample Type: | Serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids |
Assay Length: | 3.5h |
Research Area: | Cytokine;Tumor immunity;Infection immunity;Neuro science;Bone metabolism; |
Standard curve
Concentration (ng/mL) | OD | Corrected OD |
---|---|---|
2000.00 | 2.185 | 2.067 |
1000.00 | 1.687 | 1.569 |
500.00 | 1.106 | 0.988 |
250.00 | 0.826 | 0.708 |
125.00 | 0.543 | 0.425 |
62.50 | 0.354 | 0.236 |
31.25 | 0.215 | 0.097 |
0.00 | 0.118 | 0.000 |
Intra-assay Precision (Precision within an assay):CV%<8%
Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (Precision between assays):CV%<10%
Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.
Recovery
Matrices listed below were spiked with certain level of recombinant LIF and the recovery rates were calculated by comparing the measured value to the expected amount of LIF in samples.
Matrix | Recovery range | Average |
---|---|---|
serum(n=5) | 80-97% | 88% |
EDTA plasma(n=5) | 90-105% | 97% |
Heparin plasma(n=5) | 83-95% | 89% |
Linearity
The linearity of the kit was assayed by testing samples spiked with appropriate concentration of LIF and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.
Matrix | 1:2 | 1:4 | 1:8 | 1:16 |
---|---|---|---|---|
serum(n=5) | 89-98% | 83-103% | 98-105% | 92-101% |
EDTA plasma(n=5) | 92-101% | 87-98% | 85-92% | 79-96% |
Heparin plasma(n=5) | 86-95% | 85-96% | 89-100% | 85-96% |