Lablisa® Human PF4(Platelet Factor 4) ELISA Kit

Product Information

Lablisa® Human PF4(Platelet Factor 4) ELISA Kit

Catalog No : LAB1230| Pack Size : 48T , 96T

The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Human PF4. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Human PF4. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Human PF4, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Human PF4 in the samples is then determined by comparing the OD of the samples to the standard curve.

Product name:Human PF4(Platelet Factor 4) ELISA Kit
Reactivity:Human
Alternative Names:CXCL4; SCYB4; Chemokine C-X-C-Motif Ligand 4; Oncostatin-A; Iroplact
Assay Type:Sandwich
Sensitivity:0.65 ng/mL
Standard:100 ng/mL
Detection Range:1.57-100 ng/mL
Sample Type:Serum, platelet-poor plasma and cell culture supernates
Assay Length:3.5h
Research Area:Cytokine;Hematology;


Standard curve



Concentration (ng/mL)ODCorrected OD
100.002.1262.035
50.001.6281.537
25.001.2631.172
12.500.8660.775
6.250.5960.505
3.130.4810.390
1.570.2150.124
0.000.0910.000


Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.


Recovery

Matrices listed below were spiked with certain level of recombinant PF4 and the recovery rates were calculated by comparing the measured value to the expected amount of PF4 in samples.

MatrixRecovery rangeAverage
serum(n=5)78-95%86%
EDTA plasma(n=5)94-105%98%
Heparin plasma(n=5)85-97%91%


Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of PF4 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Matrix1:21:41:81:16
serum(n=5)95-104%93-105%82-101%91-103%
EDTA plasma(n=5)83-95%88-102%85-97%80-92%
Heparin plasma(n=5)86-102%87-103%85-101%81-99%