Lablisa® BH2(Dihydrobiopterin) ELISA Kit

Product Information

Lablisa® BH2(Dihydrobiopterin) ELISA Kit

Catalog No : LAB9938 | Pack Size : 48T , 96T

This assay employs the competitive inhibition enzyme immunoassay technique. The microtiter plate provided in this kit has been pre-coated with BH2. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to BH2. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of BH2 in the samples is then determined by comparing the OD of the samples to the standard curve.

Product name:Lablisa® BH2(Dihydrobiopterin) ELISA Kit
Reactivity:General
Alternative Names:Dihydrobiopterin
Assay Type:Competitive Inhibition
Sensitivity:11.31 pg/mL
Standard:2000 pg/mL
Detection Range:31.25-2000 pg/mL
Sample type:serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay length:3.5h
Research Area:Signal transduction


Standard curve



Concentration (pg/mL)ODCorrected OD
2000.000.217
1000.000.467
500.000.722
250.000.997
125.001.236
62.501.596
31.251.889
0.002.197


Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.


Recovery

Matrices listed below were spiked with certain level of recombinant BH2 and the recovery rates were calculated by comparing the measured value to the expected amount of BH2 in samples.

MatrixRecovery rangeAverage
serum(n=5)84-97%91%
EDTA plasma(n=5)82-95%88%
Heparin plasma(n=5)95-107%101%


Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of BH2 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Matrix1:21:41:81:16
serum(n=5)86-93%91-105%87-99%99-105%
EDTA plasma(n=5)80-101%87-96%96-105%95-103%
Heparin plasma(n=5)85-103%82-96%87-105%90-99%


Manuals on the web are for your reference only, specifications are subject to the delivery manuals.