Lablisa® Rat IFABP/FABP2(Intestinal Fatty Acid Binding Protein) ELISA Kit

Product Information

Lablisa® Rat IFABP/FABP2(Intestinal Fatty Acid Binding Protein) ELISA Kit

Catalog No : LAB16245 | Pack Size : 48T , 96T

The test principle applied in this kit is Sandwich enzyme immunoassay. The microtiter plate provided in this kit has been pre-coated with an antibody specific to Rat IFABP/FABP2. Standards or samples are added to the appropriate microtiter plate wells then with a biotin-conjugated antibody specific to Rat IFABP/FABP2. Next, Avidin conjugated to Horseradish Peroxidase (HRP) is added to each microplate well and incubated. After TMB substrate solution is added, only those wells that contain Rat IFABP/FABP2, biotin-conjugated antibody and enzyme-conjugated Avidin will exhibit a change in color. The enzyme-substrate reaction is terminated by the addition of sulphuric acid solution and the color change is measured spectrophotometrically at a wavelength of 450nm ± 10nm. The concentration of Rat IFABP/FABP2 in the samples is then determined by comparing the OD of the samples to the standard curve.

Product name:Lablisa® Rat IFABP/FABP2(Intestinal Fatty Acid Binding Protein) ELISA Kit
Reactivity:Rat
Alternative Names:FABPI; IFABP; I-FABP; Intestinal-type fatty acid-binding protein; FABP2; Fatty Acid Binding Protein 2, Intestinal
Assay Type:Sandwich
Sensitivity:0.139 ng/mL
Standard:20 ng/mL
Detection Range:0.32-20 ng/mL
Sample Type:serum, plasma, tissue homogenates, cell lysates, cell culture supernates and other biological fluids
Assay Length:3.5h
Research Area:Metabolic pathway;Tumor immunity;Endocrinology;Cardiovascular biology;


Standard curve



Concentration (ng/mL)ODCorrected OD
20.001.9891.900
10.001.6281.539
5.000.9760.887
2.500.7820.693
1.250.4570.368
0.630.3570.268
0.320.2360.147
0.000.0890.000


Precision

Intra-assay Precision (Precision within an assay):CV%<8%

Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.

Inter-assay Precision (Precision between assays):CV%<10%

Three samples of known concentration were tested in forty separate assays to assess inter-assay precision.


Recovery

Matrices listed below were spiked with certain level of recombinant IFABP/FABP2 and the recovery rates were calculated by comparing the measured value to the expected amount of IFABP/FABP2 in samples.

MatrixRecovery rangeAverage
serum(n=5)87-99%93%
EDTA plasma(n=5)82-95%88%
Heparin plasma(n=5)94-107%100%


Linearity

The linearity of the kit was assayed by testing samples spiked with appropriate concentration of IFABP/FABP2 and their serial dilutions. The results were demonstrated by the percentage of calculated concentration to the expected.

Matrix1:21:41:81:16
serum(n=5)88-97%96-105%81-96%93-107%
EDTA plasma(n=5)82-96%85-98%92-103%81-95%
Heparin plasma(n=5)95-102%83-96%97-105%85-97%


Manuals on the web are for your reference only, specifications are subject to the delivery manuals.